FDA Adverse Event Malfunction Summary report: N

11.0 CM LONG ATTACHMENT

MDR report key: 4203529 · Received October 27, 2014

Report

Report Number
1045834-2014-14686
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
October 9, 2014
Report Date
October 13, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE BEARINGS WERE NO LONGER IN AXIAL ALIGNMENT, WHICH DID NOT ALLOW FOR THE INSERTION OF A CUTTER. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR AND SERVICING. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WOULD NOT ALLOW THE BURRS TO SPIN. THERE WERE NO DELAYS TO THE SCHEDULED SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681619 11.0 CM LONG ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1