FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4203090 · Received September 17, 2014

Report

Report Number
2032227-2014-24357
Event Type
Malfunction
Date Received
September 17, 2014
Date of Event
August 18, 2014
Report Date
August 18, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THE INSULIN PUMP ALARMED NO DELIVERY SEVERAL TIMES. ALARM WAS RESOLVED WITH COMPLETE SET CHANGE. NO DELIVERY ALARM MIGHT INDICATE POSSIBLE OCCLUSION IN THE INFUSION SET OR THE RESERVOIR. CUSTOMER INSERTED INFUSION SET IN THE BUTTOCKS. CUSTOMER WAS ADVISED IT MIGHT BE A POSSIBLE SITE OCCLUSION OR SITE ISSUE. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
575472 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-554LWWL

Patients

Seq Age Sex Outcome Treatment
1 UNOMEDICAL INSULIN INFUSION SET