FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4203088 · Received September 18, 2014

Report

Report Number
1720753-2014-07972
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
September 2, 2014
Report Date
September 18, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE FFB AND GIB PCBS WERE EVALUATED AND REPLACED. THE CONFIGURATION FILES WERE RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE MONITOR WOULD NOT TURN ON (NO-BOOT). THE FE FOUND COMMUNICATION FAILED ERROR MESSAGES IN THE ERROR LOGS. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
578789 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1