FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 4203088
·
Received September 18, 2014
Report
- Report Number
- 1720753-2014-07972
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Date of Event
- September 2, 2014
- Report Date
- September 18, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE FFB AND GIB PCBS WERE EVALUATED AND REPLACED. THE CONFIGURATION FILES WERE RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE MONITOR WOULD NOT TURN ON (NO-BOOT). THE FE FOUND COMMUNICATION FAILED ERROR MESSAGES IN THE ERROR LOGS. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 578789 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |