HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-37556
- Event Type
- Malfunction
- Date Received
- October 27, 2014
- Date of Event
- August 17, 2014
- Report Date
- September 29, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN INCREASED INTRA-PERITONEAL VOLUME EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY THAT WAS INITIATED ON (B)(6) 2014 AT 22:35:08. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1226ML, INDICATING THE HOME PATIENT DRAINED 1226ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682018 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |