FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 4203043 · Received September 18, 2014

Report

Report Number
1720753-2014-07995
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
August 11, 2014
Report Date
September 18, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HARD DISK DRIVE WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY FAILED TO BOOT TO A USABLE STATE AND EXHIBITED INTERMITTENT FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
578827 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1