FDA Adverse Event Malfunction Summary report: N

I-STAT DOWNLOADER, NETWORK

MDR report key: 4202781 · Received October 27, 2014

Report

Report Number
2245578-2014-00077
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
October 14, 2014
Report Date
October 14, 2014
Manufacturer
ABBOTT POINT OF CARE
Product Code
JQP
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 01/12/2015. FAILURE RESULTED FROM AN UNKNOWN LIQUID INGRESS CAUSING A COMPONENT TO SHORT CIRCUIT.

Additional Manufacturer Narrative · 1

(B)(4)

Description of Event or Problem · 1

ON (B)(6) 2014, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO STATED THAT SMOKE WAS COMING FROM DOWNLOADER-PRINTER CORD/PORT OF DOWNLOADER SERIAL NUMBER (B)(4). THE PRODUCT WAS REPLACED AT NO CHARGE PER SERVICE AGREEMENT AND RETURNING FOR FURTHER INVESTIGATION. THERE WERE NO INJURIES REPORTED WITH THIS EVENT.

Description of Event or Problem · 1

NA

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
683898 I-STAT DOWNLOADER, NETWORK DOWNLOADER, NETWORK JQP ABBOTT POINT OF CARE NA NA

Patients

Seq Age Sex Outcome Treatment
1