Malfunction
Summary report: N
I-STAT DOWNLOADER, NETWORK
MDR report key: 4202781
·
Received October 27, 2014
Report
- Report Number
- 2245578-2014-00077
- Event Type
- Malfunction
- Date Received
- October 27, 2014
- Date of Event
- October 14, 2014
- Report Date
- October 14, 2014
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JQP
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE INVESTIGATION WAS COMPLETED ON 01/12/2015. FAILURE RESULTED FROM AN UNKNOWN LIQUID INGRESS CAUSING A COMPONENT TO SHORT CIRCUIT.
Additional Manufacturer Narrative · 1
(B)(4)
Description of Event or Problem · 1
ON (B)(6) 2014, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO STATED THAT SMOKE WAS COMING FROM DOWNLOADER-PRINTER CORD/PORT OF DOWNLOADER SERIAL NUMBER (B)(4). THE PRODUCT WAS REPLACED AT NO CHARGE PER SERVICE AGREEMENT AND RETURNING FOR FURTHER INVESTIGATION. THERE WERE NO INJURIES REPORTED WITH THIS EVENT.
Description of Event or Problem · 1
NA
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 683898 | I-STAT DOWNLOADER, NETWORK | DOWNLOADER, NETWORK | JQP | ABBOTT POINT OF CARE | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |