FDA Adverse Event Malfunction Summary report: N

XLD 110V LATIN AMERI

MDR report key: 4202588 · Received September 18, 2014

Report

Report Number
9615050-2014-05304
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
January 1, 2014
Report Date
September 14, 2014
Manufacturer
HOSPIRA COSTA RICA LTD
Product Code
FRN
PMA / PMN Number
K010924
Removal / Correction Number
FA311-04
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER PIN WAS MISSING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER PIN WAS MISSING. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A NOTE THAT STATED DAMAGED. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS, DELAYS IN CRITICAL THERAPIES, OR NECESSITY FOR MEDICAL INTERVENTIONS WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
580760 XLD 110V LATIN AMERI 80FRN FRN HOSPIRA COSTA RICA LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK