FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4202574 · Received October 27, 2014

Report

Report Number
1416980-2014-37521
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
October 1, 2014
Report Date
October 1, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: CMPLNT-(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, NOT PRIMING THE LINE COMPLETELY IS A KNOWN CAUSE OF THE REPORTED ALARM. PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. THE PATIENT IS ALSO TOLD TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED BEFORE CONNECTING. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR, AIR IN SET/LINE ALARM OCCURRED ON A HOME CHOICE (HC) DEVICE DURING INITIAL DRAIN. THE PATIENT WAS CONNECTED TO THE HOMECHOICE AT THE TIME OF THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE EXPLAINED THE ALARM AND ASSISTED THE PATIENT IN CLEARING THE ALARM. DURING THE TROUBLESHOOTING, IT WAS DETERMINED THAT THE PATIENT CONNECTED TO THE DEVICE BEFORE COMPLETELY PRIMING THE PATIENT LINE. THE PATIENT WAS TO START THERAPY OVER WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
684669 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 42 YR HOMECHOICE