FDA Adverse Event Injury Summary report: N

EON

MDR report key: 4202017 · Received October 17, 2014

Report

Report Number
1627487-2014-21671
Event Type
Injury
Date Received
October 17, 2014
Date of Event
January 25, 2014
Report Date
September 25, 2014
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 2. REF MFR REPORT: 1627487-2014-21672. THE PT REPORTED EXPERIENCING POCKET HEATING AT THE IPG SITE WHILE CHARGING. A REPLACEMENT CHARGER WAS SENT TO THE PT WHICH RESOLVED THE ISSUE. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663314 EON SCS IPG GZB ST. JUDE MEDICAL - NEUROMODULATION 3716 96907

Patients

Seq Age Sex Outcome Treatment
1 49 YR Other IMPLANT DATE:| SCS LEAD, MODEL 3210,