EON
Report
- Report Number
- 1627487-2014-23687
- Event Type
- Injury
- Date Received
- October 17, 2014
- Date of Event
- May 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- ST. JUDE MEDICAL ¿ NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE #1 OF 3. REFERENCE MFR. REPORT: 1627487-2014-23688. THE PATIENT REPORTED SHE EXPERIENCED POCKET HEATING WHILE RECHARGING HER IPG. A REPLACEMENT CHARGING SYSTEM WAS SENT TO THE PATIENT TO ADDRESS THE ISSUE. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 663342 | EON | SCS IPG | GZB | ST. JUDE MEDICAL ¿ NEUROMODULATION | 3716 | 119789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Other | SCS LEAD: MODEL 3214| IMPLANT DATE: |