FDA Adverse Event Malfunction Summary report: N

TENDRIL STS

MDR report key: 4201620 · Received January 13, 2014

Report

Report Number
2017865-2014-04719
Event Type
Malfunction
Date Received
January 13, 2014
Date of Event
November 25, 2011
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DTB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION DESCRIPTION: TEST PERFORMED INDICATED NORMAL ELECTRICAL CHARACTERISTICS. VISUAL INSPECTION DI NOT FIND ANY ANOMALIES BEYOND HELIX BEING CLOGGED WITH DRIED BLOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED NOISE. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29577 TENDRIL STS PERMANENT PACEMAKER ELECTRODE DTB ST. JUDE MEDICAL, INC., CRMD 2088TC/52

Patients

Seq Age Sex Outcome Treatment
1 73 YR