GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00977
- Event Type
- Malfunction
- Date Received
- October 27, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 18, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER 0010-2400-415A-(B)(4): THE FIBER/GLASS CAP EXHIBITED A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE ROTATED INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITED MILD DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITED MILD BURNT ON DETRITUS. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY ALSO EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/ PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER. (B)(4) REFERS TO FORWARD FIRING OF THE SIDE FIRING SURGICAL FIBER; A CODE REQUEST HAS BEEN SUBMITTED.
IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, "THE TIP OF THE LASER BURNED OFF AFTER BEING USED ON PROSTATE STONES". THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. THERE WAS ¿NO INJURY TO PATIENT¿ REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 681283 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 415A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |