FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 4201618 · Received October 27, 2014

Report

Report Number
2937094-2014-00977
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
September 8, 2014
Report Date
September 18, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER 0010-2400-415A-(B)(4): THE FIBER/GLASS CAP EXHIBITED A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE ROTATED INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITED MILD DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITED MILD BURNT ON DETRITUS. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY ALSO EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/ PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER. (B)(4) REFERS TO FORWARD FIRING OF THE SIDE FIRING SURGICAL FIBER; A CODE REQUEST HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, "THE TIP OF THE LASER BURNED OFF AFTER BEING USED ON PROSTATE STONES". THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. THERE WAS ¿NO INJURY TO PATIENT¿ REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681283 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 415A

Patients

Seq Age Sex Outcome Treatment
1