FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4201319 · Received October 27, 2014

Report

Report Number
2032227-2014-43014
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
September 26, 2014
Report Date
September 27, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED PRIME DURING THE BASIC OCCLUSION TEST, DUE TO PROTRUDED AND LOOSE DRIVE SUPPORT DISK AND UNABLE TO PRIME DURING THE PRIME TEST. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Description of Event or Problem · 1

CUSTOMER REPORTED A COMPROMISED FORCE SENSOR SYSTEM ALARM WHILE PRIMING. CUSTOMER'S BLOOD GLUCOSE WAS 200 MG/DL. TROUBLESHOOTING FOUND THE CUSTOMER'S DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.