PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-43014
- Event Type
- Malfunction
- Date Received
- October 27, 2014
- Date of Event
- September 26, 2014
- Report Date
- September 27, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP ALARMED PRIME DURING THE BASIC OCCLUSION TEST, DUE TO PROTRUDED AND LOOSE DRIVE SUPPORT DISK AND UNABLE TO PRIME DURING THE PRIME TEST. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)
CUSTOMER REPORTED A COMPROMISED FORCE SENSOR SYSTEM ALARM WHILE PRIMING. CUSTOMER'S BLOOD GLUCOSE WAS 200 MG/DL. TROUBLESHOOTING FOUND THE CUSTOMER'S DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 683810 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |