FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 4201237 · Received October 27, 2014

Report

Report Number
2955842-2014-05541
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
September 2, 2014
Report Date
September 24, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT THE PITCH CABLE IS BROKEN AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DOES NOT EXHIBIT ANY WEAR. THE BROKEN STRANDS STICK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, BROKEN PITCH CABLE, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR STARTING A DA VINCI SURGICAL PROCEDURE, IT WAS NOTED THAT THE CABLE ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS BROKEN. THERE WAS NO REPORT THAT ANY PIECE(S) FROM THE INSTRUMENT FELL INTO A PATIENT AND THERE WAS NO ALLEGATION OF ANY PATIENT HARM, ADVERSE OUTCOME OR INJURY INVOLVING THE REPORTED INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681938 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-07 N10140606 439

Patients

Seq Age Sex Outcome Treatment
1