FENESTRATED BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-05541
- Event Type
- Malfunction
- Date Received
- October 27, 2014
- Date of Event
- September 2, 2014
- Report Date
- September 24, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT THE PITCH CABLE IS BROKEN AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DOES NOT EXHIBIT ANY WEAR. THE BROKEN STRANDS STICK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, BROKEN PITCH CABLE, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT PRIOR STARTING A DA VINCI SURGICAL PROCEDURE, IT WAS NOTED THAT THE CABLE ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS BROKEN. THERE WAS NO REPORT THAT ANY PIECE(S) FROM THE INSTRUMENT FELL INTO A PATIENT AND THERE WAS NO ALLEGATION OF ANY PATIENT HARM, ADVERSE OUTCOME OR INJURY INVOLVING THE REPORTED INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 681938 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-07 | N10140606 439 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |