FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4201197 · Received October 27, 2014

Report

Report Number
2032227-2014-42811
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
September 25, 2014
Report Date
September 25, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH AN INTERMITTENT BUTTONS, DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO TRACES OF MOISTURE WERE NOTED AT ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. INSULIN PUMP HAD A CRACKED CASE AT DISPLAY WINDOW CORNERS, DISPLAY WINDOW AND BATTERY TUBE THREADS AND MINOR SCRATCHES ON LCD WINDOW.

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS RECEIVING A BUTTON ERROR ALARM ON HIS INSULIN PUMP. HIS BLOOD GLUCOSE WAS 61MG/DL AT THE TIME OF REPORTING. THE CUSTOMER WAS ADVISED TO REVERT TO HIS BACKUP PLAN. THE PUMP IS BEING RETURNED. NO ADDITIONAL INFORMATION TO RETURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681657 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 21 YR