FDA Adverse Event Malfunction Summary report: N

EASYLINK

MDR report key: 4201189 · Received October 27, 2014

Report

Report Number
2517506-2014-00243
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
October 1, 2014
Report Date
October 1, 2014
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JQP
PMA / PMN Number
K051087
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE CSE ANALYZED THE INSTRUMENT AND DETERMINED THAT THE CAUSE OF THE SMOKE WAS DUE TO A MALFUNCTION IN THE INSTRUMENT UNINTERRUPTIBLE POWER SUPPLY (UPS). THE CSE REPLACED THE UPS AND THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. FURTHER EVALUATION OF THE DEVICE IS NOT REQUIRED.

Description of Event or Problem · 1

SMOKE EMISSION WAS DETECTED ON A EASYLINK SYSTEM UNINTERRUPTIBLE POWER SUPPLY (UPS). THE UPS POWER WAS TURNED OFF TO STOP THE SMOKE EMISSION. THERE WERE NO REPORTS OF STAFF INJURY, DAMAGE TO PROPERTY OR IMPACT TO PATIENTS DUE TO THE SMOKE EMISSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682072 EASYLINK INFORMATICS SYSTEM JQP SIEMENS HEALTHCARE DIAGNOSTICS INC. EASYLINK INFORMATICS SYSTEM

Patients

Seq Age Sex Outcome Treatment
1