FDA Adverse Event
Malfunction
Summary report: N
EASYLINK
MDR report key: 4201189
·
Received October 27, 2014
Report
- Report Number
- 2517506-2014-00243
- Event Type
- Malfunction
- Date Received
- October 27, 2014
- Date of Event
- October 1, 2014
- Report Date
- October 1, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JQP
- PMA / PMN Number
- K051087
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE CSE ANALYZED THE INSTRUMENT AND DETERMINED THAT THE CAUSE OF THE SMOKE WAS DUE TO A MALFUNCTION IN THE INSTRUMENT UNINTERRUPTIBLE POWER SUPPLY (UPS). THE CSE REPLACED THE UPS AND THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. FURTHER EVALUATION OF THE DEVICE IS NOT REQUIRED.
Description of Event or Problem · 1
SMOKE EMISSION WAS DETECTED ON A EASYLINK SYSTEM UNINTERRUPTIBLE POWER SUPPLY (UPS). THE UPS POWER WAS TURNED OFF TO STOP THE SMOKE EMISSION. THERE WERE NO REPORTS OF STAFF INJURY, DAMAGE TO PROPERTY OR IMPACT TO PATIENTS DUE TO THE SMOKE EMISSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682072 | EASYLINK | INFORMATICS SYSTEM | JQP | SIEMENS HEALTHCARE DIAGNOSTICS INC. | EASYLINK INFORMATICS SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |