FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4201140 · Received October 27, 2014

Report

Report Number
2032227-2014-42886
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
September 26, 2014
Report Date
September 26, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED BASIC OCCLUSION TEST AND DISPLACEMENT TEST. NO MOTOR ERROR ALARMS WERE NOTED. HOWEVER, THE DEVICE WAS UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTOR. NO COMPROMISED FORCE SENSOR SYSTEM ALARMS WERE NOTED. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. THE DEVICE HAD MINOR SCRATCHED LCD WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THERE IS A MOTOR ERROR ALARM AND ANOTHER ALARM. THE BLOOD GLUCOSE READING IS 95 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682548 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 44 YR