FDA Adverse Event Malfunction Summary report: N

ACCESS 2 IMMUNOASSAY SYSTEM

MDR report key: 4201091 · Received October 27, 2014

Report

Report Number
2122870-2014-00745
Event Type
Malfunction
Date Received
October 27, 2014
Date of Event
October 6, 2014
Report Date
October 6, 2014
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K922823/A007
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE SYSTEM AND IDENTIFIED A FLUID LEAK AT THE SUBSTRATE FLUIDICS LINE DUE TO A LOOSE FITTING. THE FSE TIGHTENED THE LOOSE FITTING AND RESOLVED THE ISSUE. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. IN CONCLUSION, THE LIKELY CAUSE OF THE EVENT IS ATTRIBUTED TO THE SUBSTRATE LINE FITTING. A DEFINITIVE ROOT CAUSE IS UNKNOWN.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SUBSTRATE BOTTLE EMPTY WITH TESTS REMAINING AND SAMPLE COUNTS OUTSIDE LIMITS SYSTEM ERRORS INVOLVING THE ACCESS 2 IMMUNOASSAY SYSTEM. THE CUSTOMER STATED THERE WERE NO ERRONEOUS PATIENT RESULTS GENERATED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682092 ACCESS 2 IMMUNOASSAY SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1