FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 4200883 · Received September 17, 2014

Report

Report Number
9615739-2014-00358
Event Type
Malfunction
Date Received
September 17, 2014
Date of Event
August 20, 2014
Report Date
August 20, 2014
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

HILL-ROM TECH SUPPORT RECOMMENDED CHANGING THE SCALE CONTROL BOARD. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE ACCOUNT REPLACED THE SCALE CONTROL BOARD TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED EXIT ALARM WAS NOT ALARMING AT THE BED. THE BED WAS LOCATED IN THE MED SURG DEPT. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).