M2A 38MM MOD HD +3MM NK
Report
- Report Number
- 0001825034-2014-08327
- Event Type
- Injury
- Date Received
- October 24, 2014
- Date of Event
- January 18, 2012
- Report Date
- April 13, 2015
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWA
- PMA / PMN Number
- PK011110
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- ATTORNEY
Narratives
THE FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION THAT WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.
THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION THAT WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, IT STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ AND ¿INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS.¿ AND ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ AND ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS NUMBER 1 OF 2 MDR'S FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-08327 & -09311).
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. EVENT IS BEING REPORTED TO FDA ON ONE MEDWATCH AS THE LIMITED INFORMATION AVAILABLE INDICATES THAT A REVISION PROCEDURE OCCURRED. SHOULD ADDITIONAL INFORMATION BE RECEIVED REGARDING THE REVISION PROCEDURE, THE COMPLAINT WILL BE REASSESSED AND FURTHER MEDWATCH REPORTS WILL BE SUBMITTED, IF NECESSARY.
PATIENT'S LEGAL COUNSEL REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2004. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, METAL POISONING, FATIGUE, AND AN AFFECTED GAIT. REVIEW OF INVOICE RECORDS INDICATES THE CUP AND MODULAR HEAD WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED IN OPERATIVE REPORT NOTED PATIENT UNDERWENT A RIGHT HIP REVISION ON (B)(6) 2012 DUE TO METAL ON METAL WEAR WITH OSTEOLYSIS. OPERATIVE REPORT NOTED THE PRESENCE OF REACTIVE SYNOVITIS, CREAMY EXUDATE CONSISTENT WITH METALLOSIS, METAL ON METAL WEAR, CORROSION ON THE TAPER AND CAVITARY LESIONS IN THE ANTERIOR INFERIOR ASPECT OF THE ACETABULUM. THE MODULAR HEAD AND ACETABULAR CUP WERE REMOVED AND REPLACED.
PATIENT'S LEGAL COUNSEL REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2004. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, METAL POISONING, FATIGUE, AND AN AFFECTED GAIT. REVIEW OF INVOICE RECORDS INDICATES THE CUP AND MODULAR HEAD WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
PATIENT'S LEGAL COUNSEL REPORTED PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6) 2004. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, METAL POISONING, FATIGUE, AND AN AFFECTED GAIT. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 679315 | M2A 38MM MOD HD +3MM NK | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 179590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Hospitalization| R |