FDA Adverse Event Injury Summary report: N

M2A 38MM MOD HD +3MM NK

MDR report key: 4200695 · Received October 24, 2014

Report

Report Number
0001825034-2014-08327
Event Type
Injury
Date Received
October 24, 2014
Date of Event
January 18, 2012
Report Date
April 13, 2015
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK011110
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

THE FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION THAT WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION THAT WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, IT STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ AND ¿INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS.¿ AND ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ AND ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS NUMBER 1 OF 2 MDR'S FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-08327 & -09311).

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. EVENT IS BEING REPORTED TO FDA ON ONE MEDWATCH AS THE LIMITED INFORMATION AVAILABLE INDICATES THAT A REVISION PROCEDURE OCCURRED. SHOULD ADDITIONAL INFORMATION BE RECEIVED REGARDING THE REVISION PROCEDURE, THE COMPLAINT WILL BE REASSESSED AND FURTHER MEDWATCH REPORTS WILL BE SUBMITTED, IF NECESSARY.

Description of Event or Problem · 1

PATIENT'S LEGAL COUNSEL REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2004. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, METAL POISONING, FATIGUE, AND AN AFFECTED GAIT. REVIEW OF INVOICE RECORDS INDICATES THE CUP AND MODULAR HEAD WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED IN OPERATIVE REPORT NOTED PATIENT UNDERWENT A RIGHT HIP REVISION ON (B)(6) 2012 DUE TO METAL ON METAL WEAR WITH OSTEOLYSIS. OPERATIVE REPORT NOTED THE PRESENCE OF REACTIVE SYNOVITIS, CREAMY EXUDATE CONSISTENT WITH METALLOSIS, METAL ON METAL WEAR, CORROSION ON THE TAPER AND CAVITARY LESIONS IN THE ANTERIOR INFERIOR ASPECT OF THE ACETABULUM. THE MODULAR HEAD AND ACETABULAR CUP WERE REMOVED AND REPLACED.

Description of Event or Problem · 1

PATIENT'S LEGAL COUNSEL REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2004. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, METAL POISONING, FATIGUE, AND AN AFFECTED GAIT. REVIEW OF INVOICE RECORDS INDICATES THE CUP AND MODULAR HEAD WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

PATIENT'S LEGAL COUNSEL REPORTED PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6) 2004. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, METAL POISONING, FATIGUE, AND AN AFFECTED GAIT. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
679315 M2A 38MM MOD HD +3MM NK PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 179590

Patients

Seq Age Sex Outcome Treatment
1 38 YR Hospitalization| R