FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4200599 · Received October 24, 2014

Report

Report Number
2032227-2014-42663
Event Type
Malfunction
Date Received
October 24, 2014
Date of Event
September 24, 2014
Report Date
September 24, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED DISPLACEMENT TEST AND REWIND TEST. THE DEVICE ALARMED DURING BASIC OCCLUSION TEST DUE TO LOOSE AND OR PROTRUDED DRIVE SUPPORT DISK. THE DEVICE HAD MINOR SCRATCHES ON LCD WINDOW, BROKEN RESERVOIR TUBE LIP, SCRATCHES ON RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT, AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING AN ERROR ALARM DURING THE MANUAL PRIME AND STATED THAT INSULIN WAS SQUIRTING OUT. CUSTOMER STATED THAT THE PISTON WAS NOT STOPPING DURING THE FULL TUBING. THE DRIVE SUPPORT CAP WAS NOTED TO BE PROTRUDED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 188 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678618 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 22 YR