FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4200598 · Received October 24, 2014

Report

Report Number
2032227-2014-42660
Event Type
Malfunction
Date Received
October 24, 2014
Date of Event
September 24, 2014
Report Date
September 24, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO FLATTENED BUTTON DOME SWITCH ON ALL BUTTONS. THE DEVICE HAD MINOR SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT AT BATTERY TUBE THREADS AREA, AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP FREQUENTLY HAD NO OR INTERMITTENT RESPONSE BY BUTTON PRESS. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 82 MG/DL. NO FURTHER INFORMATION REPORTED.