Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4200594
·
Received October 24, 2014
Report
- Report Number
- 2032227-2014-42669
- Event Type
- Malfunction
- Date Received
- October 24, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS WERE NOTED. THE INSULIN PUMP HAD CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY AND RESERVOIR WINDOW, AND A MISSING END CAP STICKER.
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 124 MG/DL. NO FURTHER INFORMATION REPORTED.