FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4200590 · Received October 24, 2014

Report

Report Number
2032227-2014-42661
Event Type
Malfunction
Date Received
October 24, 2014
Date of Event
October 16, 2014
Report Date
October 16, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4). PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO FLATTENED ACT BUTTON DOME SWITCH. NO BUTTON ERROR ALARM DURING TESTING. PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND EXCESSIVE NO DELIVERY TEST. NO COSMETIC DAMAGE NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER STATED THAT HE HAD BLOOD GLUCOSE READINGS OF 383 MG/DL WHICH IS NOT USUAL FOR HIM, HOWEVER HAS TREATED AND FEELS OK. CUSTOMER'S MOST RECENT BLOOD GLUCOSE READING WAS 73 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678615 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 14 YR