FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4200561
·
Received October 24, 2014
Report
- Report Number
- 2032227-2014-42728
- Event Type
- Malfunction
- Date Received
- October 24, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
INSULIN PUMP ALARMED PRIME DURING BASIC OCCLUSION TEST, DUE TO LOOSE AND PROTRUDED DRIVE SUPPORT DISK. INSULIN PUMP RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW AND CRACKED BATTERY TUBE THREADS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING PRIME AND INSULIN WAS SQUIRT OUT OF THE TUBING. THE INSULIN PUMP WAS DROPPED PRIOR TO THE ALARM. THE DRIVE SUPPORT CAP IS PROTRUDED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 679139 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |