FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 4200497 · Received October 24, 2014

Report

Report Number
2531779-2014-30351
Event Type
Malfunction
Date Received
October 24, 2014
Report Date
October 8, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/17/2014 WITH THE FOLLOWING FINDINGS: THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED BELOW THE BUMPER PAD. UNRELATED TO THE INITIAL COMPLAINT, THE DISPLAY WAS FOUND TO BE DIM AND FADING WITH A REDDISH TINT. THE UP, DOWN, AND OK BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE TO TESTING. THE KEYPAD COVER WAS FOUND TO BE INTACT; THE KEYPAD COVER WAS REMOVED FOR FURTHER INVESTIGATION AND CONTAMINATION WAS FOUND UNDER ALL THE KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER STATED THAT THE DISPLAY WAS DIM AND DENIED MOISTURE OR CORROSION IN THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678846 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 49 YR