ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2014-30351
- Event Type
- Malfunction
- Date Received
- October 24, 2014
- Report Date
- October 8, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/17/2014 WITH THE FOLLOWING FINDINGS: THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED BELOW THE BUMPER PAD. UNRELATED TO THE INITIAL COMPLAINT, THE DISPLAY WAS FOUND TO BE DIM AND FADING WITH A REDDISH TINT. THE UP, DOWN, AND OK BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE TO TESTING. THE KEYPAD COVER WAS FOUND TO BE INTACT; THE KEYPAD COVER WAS REMOVED FOR FURTHER INVESTIGATION AND CONTAMINATION WAS FOUND UNDER ALL THE KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER STATED THAT THE DISPLAY WAS DIM AND DENIED MOISTURE OR CORROSION IN THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.