UNICEL DXC 800 PRO SYNCHRON SYSTEM
Report
- Report Number
- 2050012-2014-00478
- Event Type
- Injury
- Date Received
- October 24, 2014
- Date of Event
- October 1, 2014
- Report Date
- October 1, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- CGA
- PMA / PMN Number
- K103842
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND EVALUATED THE INSTRUMENT. THE FSE FOUND THE ACCUSENSE GLUCOSE OXYGEN SENSOR ELECTRODE DAMAGED CAUSED BY CUSTOMER USE ERROR. THE CUSTOMER HAD INCORRECTLY INSTALLED THE ACCUSENSE GLUCOSE OXYGEN SENSOR ELECTRODE DURING MAINTENANCE. THE FSE REPLACED THE ACCUSENSE GLUCOSE OXYGEN SENSOR ELECTRODE TO RESOLVE THE ISSUE.
THE CUSTOMER REPORTED OBTAINING AN ERRONEOUSLY LOW GLUM (GLUCOSE) RESULT FOR ONE (1) PATIENT, INVOLVING THE UNICEL DXC 800 PRO SYNCHRON SYSTEM. THE RESULT WAS QUESTIONED WHEN THERE WAS A DISCREPANCY BETWEEN THE UNICEL DXC 800 PRO SYNCHRON SYSTEM RESULT AND AN ALTERNATE METHODOLOGY. THE CUSTOMER REPEATED THE SAMPLE ON ANOTHER UNICEL DXC SYSTEM AND OBTAINED A HIGHER GLUM RESULT. THE ERRONEOUSLY LOW GLUM RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE PATIENT RECEIVED AN IV (INTRAVENOUS) DUE TO THE ERRONEOUSLY LOW GLUM RESULT. QUALITY CONTROL WAS WITHIN LABORATORY'S ESTABLISHED RANGES ON THE DAY OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 678949 | UNICEL DXC 800 PRO SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | CGA | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |