FDA Adverse Event Malfunction Summary report: N

VISAO®

MDR report key: 4200437 · Received October 24, 2014

Report

Report Number
1045254-2014-00269
Event Type
Malfunction
Date Received
October 24, 2014
Date of Event
September 9, 2014
Report Date
September 29, 2014
Manufacturer
MEDTRONIC XOMED INC.
Product Code
ERL
PMA / PMN Number
K011321
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. (B)(4). THE PRODUCT INSPECTION DETERMINED THAT THIS HANDPIECE GOT HOT AFTER 30 SECONDS. THE NOSE'S TEMPERATURE WAS 57.8 DEGREES C, THE MIDDLE WAS 25.1 DEGREES C, AND THE REAR WAS 25 DEGREES C. IT WAS ALSO NOISY WHEN RUNNING. THE FULL EVALUATION HAS NOT BEEN COMPLETED AT THIS TIME. METHOD: TEMPERATURE TESTING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OTOLOGIC SURGERY, THE HANDPIECE GOT HOT TO THE TOUCH WITHIN 1 MINUTE. THE PROCEDURE WAS COMPLETED WITH ANOTHER HANDPIECE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678845 VISAO® DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL MEDTRONIC XOMED INC. 3334800 205317262

Patients

Seq Age Sex Outcome Treatment
1