HARMONIC ACE ADAPTIVE TISSUE TECHNOLOGY
Report
- Report Number
- 3005075853-2014-07350
- Event Type
- Malfunction
- Date Received
- October 24, 2014
- Date of Event
- August 21, 2014
- Report Date
- August 26, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K120729
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE RECEIVED WAS A HAR36 (DEVICE USED TO COMPLETE THE PROCEDURE) AND NOT THE HARH36 WHICH WAS THE DEVICE USED DURING THE PROCEDURE WITH THE REPORTED PROBLEM OF REPLACE INSTRUMENT ERROR. DURING FUNCTIONAL TESTING ON A GENERATOR THE DEVICE GAVE AN ALERT SCREEN. THE DEVICE WILL STOP ACTIVATING, AND DISPLAY AN ALERT SCREEN ON THE GENERATOR WHEN THE BLADE BECOMES DAMAGED. THE DEVICE WAS DISASSEMBLED AND THE BREAK WAS LOCATED INSIDE THE TUBE PROXIMAL TO THE TISSUE PAD. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿ OR ¿BLADE ERROR DETECTED¿ "RELAXED PRESSURE IN BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. THE ANALYSIS CONCLUDED THAT THE BLADE BROKE OFF DUE TO CONTACT WITH THE CLAMP ARM PIN. NO CONCLUSION COULD BE REACHED AS TO WHAT CAUSED THE BLADE TO MAKE CONTACT WITH THE CLAMP ARM PIN.
IT WAS REPORTED THAT DURING A LAP. HYSTERECTOMY PROCEDURE, THE GENERATOR DISPLAYED 'REPLACE INSTRUMENT" WHILE USING A HARH36. THE PROCEDURE WAS COMPLETED USING A HAR36 AND THERE WERE NO CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 679710 | HARMONIC ACE ADAPTIVE TISSUE TECHNOLOGY | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | L91J8E |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |