PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE
Report
- Report Number
- 1419937-2014-00795
- Event Type
- Malfunction
- Date Received
- October 21, 2014
- Date of Event
- October 15, 2014
- Report Date
- October 17, 2014
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
PRODUCT WAS RECEIVED ON (B)(4) 2014 AND EVALUATED ON (B)(4) 2014 BY QE AND A BREACH IN THE TRANSFORMER HOUSING WAS DISCOVERED. THE CAUSE OF THE BREACH IN THE REV N TRANSFORMER HAS NOT BEEN DETERMINED AT THIS TIME. IT IS CURRENTLY BEING INVESTIGATED UNDER (B)(4). A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS BREACHED, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS NOT PLUGGED IN SO THE VOLTAGE ACROSS THE DC PLUG WAS NOT MEASURED. RESISTANCE ACROSS THE DC PLUG WAS MEASURED AT 0.770 M OHMS, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 54.8, WHICH IS NOT NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.
THE CUSTOMER CALLED IN TO ORIGINALLY REPORT THAT HER TRANSFORMER HOUSING HAD A CRACK. UPON RECEIPT OF THE PRODUCT AND A QE EVALUATION ON (B)(4) 2014, A BREACH IN THE TRANSFORMER HOUSING WAS OBSERVED WHICH IS A SAFETY RISK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 670237 | PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE | HGX | MEDELA, INC. | 57063/9207010 | REV N-1313 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |