FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

MDR report key: 4200155 · Received October 21, 2014

Report

Report Number
1419937-2014-00795
Event Type
Malfunction
Date Received
October 21, 2014
Date of Event
October 15, 2014
Report Date
October 17, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS RECEIVED ON (B)(4) 2014 AND EVALUATED ON (B)(4) 2014 BY QE AND A BREACH IN THE TRANSFORMER HOUSING WAS DISCOVERED. THE CAUSE OF THE BREACH IN THE REV N TRANSFORMER HAS NOT BEEN DETERMINED AT THIS TIME. IT IS CURRENTLY BEING INVESTIGATED UNDER (B)(4). A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS BREACHED, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS NOT PLUGGED IN SO THE VOLTAGE ACROSS THE DC PLUG WAS NOT MEASURED. RESISTANCE ACROSS THE DC PLUG WAS MEASURED AT 0.770 M OHMS, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 54.8, WHICH IS NOT NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.

Description of Event or Problem · 1

THE CUSTOMER CALLED IN TO ORIGINALLY REPORT THAT HER TRANSFORMER HOUSING HAD A CRACK. UPON RECEIPT OF THE PRODUCT AND A QE EVALUATION ON (B)(4) 2014, A BREACH IN THE TRANSFORMER HOUSING WAS OBSERVED WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
670237 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC. 57063/9207010 REV N-1313

Patients

Seq Age Sex Outcome Treatment
1