DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2014-00523
- Event Type
- Malfunction
- Date Received
- October 24, 2014
- Date of Event
- September 29, 2014
- Report Date
- September 30, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JGS
- PMA / PMN Number
- K051087
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SIEMENS HEADQUARTERS SUPPORT CENTER (HSC) SPECIALIST EVALUATED THE INSTRUMENT DATA, WHICH INDICATED THAT THE INTEGRATED MULTISENSOR TECHNOLOGY (IMT) SYSTEM DILUTION OF THE SAMPLE WAS IMPACTED BY PARTICULATE MATTER PRESENT IN THE SAMPLE. THE SAMPLE HAD ARRIVED PRE-CENTRIFUGED TO THE LAB AND THE SAMPLE HANDLING DETAILS ARE UNKNOWN. THERE WERE NO ADDITIONAL DISCORDANT ELECTROLYTE RESULTS FOR OTHER PATIENT SAMPLES. THE CAUSE OF THE DISCORDANT NA, K, AND CL RESULTS ON ONE PATIENT SAMPLE IS RELATED TO POOR SAMPLE QUALITY. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
DISCORDANT, FALSELY LOW RESULTS WERE OBTAINED ON ONE PATIENT SAMPLE TESTED FOR SODIUM (NA), POTASSIUM (K), AND CHLORIDE (CL) ON A DIMENSION VISTA 1500 INSTRUMENT. THE DISCORDANT RESULTS WERE REPORTED TO THE PHYSICIAN(S), WHO SENT THE PATIENT TO AN EMERGENCY ROOM (ER). THE PATIENT WAS REDRAWN AT THE ER AND THE NEW SAMPLE WAS TESTED AT THE ER CLINIC LAB, RESULTING HIGHER FOR NA. THE K AND CL RESULTS FROM THE SAMPLE OBTAINED IN THE ER ARE UNKNOWN. THE LABORATORY THEN REPEATED THE INITIAL SAMPLE ON AN ALTERNATE DIMENSION VISTA INSTRUMENT AND THE NA, K, AND CL RESULTS WERE ALL HIGHER. THE CORRECTED RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT NA, K, AND CL RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 678739 | DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY ANALYZER | JGS | SIEMENS HEALTHCARE DIAGNOSTICS INC. | DIMENSION VISTA 1500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |