FDA Adverse Event Injury Summary report: N

28Z AMBULANCE COT

MDR report key: 4193808 · Received October 8, 2014

Report

Report Number
1523574-2014-00020
Event Type
Injury
Date Received
October 8, 2014
Date of Event
September 4, 2014
Report Date
October 7, 2014
Manufacturer
FERNO-WASHINGTON, INC.
Product Code
FPO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COT WAS EVALUATED AND TESTED BY EMSAR, ON BEHALF OF THE MFR, ON (B)(6) 2014. THE UNIT WAS WEIGHT TESTED AND THE INCIDENT COULD NOT BE DUPLICATED. NO VISUAL MALFUNCTIONS WERE OBSERVED.

Description of Event or Problem · 1

CREW WAS PULLING COT FROM AMBULANCE. THEY SAW THE WHEELS GO DOWN BUT WERE UNSURE OF HEARING WHEELS LOCK. THEY STARTED TO PULL COT OUT AND THE COT AND PT LOWERED TO THE GROUND. PT HIT SIDE OF HEAD ON THE GROUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632777 28Z AMBULANCE COT 28Z AMBULANCE COT FPO FERNO-WASHINGTON, INC. 0012805

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention