FDA Adverse Event Injury Summary report: N

HAKIM INLINE PROGRAMMABLE VALVE SG

MDR report key: 4193619 · Received October 22, 2014

Report

Report Number
1226348-2014-12067
Event Type
Injury
Date Received
October 22, 2014
Date of Event
August 30, 2014
Manufacturer
CODMAN & SHURTLEFF, INC
Product Code
JXG
PMA / PMN Number
PK992173
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 150MMH2O. THE VALVE WAS VISUALLY INSPECTED: NO DEFECTS WERE NOTED. THE VALVE WAS IRRIGATED WITH PURIFIED WATER. NO OCCLUSION WAS NOTED. THE SIPHON GUARD WAS IRRIGATED WITH PURIFIED WATER, NO OCCLUSION WAS NOTED. THE VALVE WAS DRIED. THE VALVE WAS LEAK TESTED, NO LEAKS WERE NOTED. THE VALVE WAS TESTED FOR PROGRAMMING. THE VALVE PASSED THE TEST. THE SIPHON GUARD WAS REMOVED FOR THE PRESSURE TEST. THE VALVE WAS THEN PRESSURE TESTED, THE VALVE PASSED THE TEST. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE, PRODUCT CODE 82-3162, WITH LOT NUMBER CRGCCD, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK ON THE 19TH JUNE 2014. NO ROOT CAUSE COULD BE DETERMINED AS NO PROBLEM WAS FOUND WITH THE VALVES. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CONSIDERED CLOSED.

Description of Event or Problem · 1

SALES REP. DOESN'T HAVE ANY DETAILS AT THIS TIME. HE IS WAITING TO GET MORE INFORMATION FROM THE DOCTOR REGARDING THE DETAILS OF THE PROBLEM. ON 10/22/2014, THE REP HAS MADE A COUPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION ABOUT THIS COMPLAINT. HE HAS COMMUNICATED THAT THE HOSPITAL HAS NOT BEEN RESPONSIVE. AS A RESULT THE COMPLAINT WILL BE FILED AS IT APPEARS AN EXPLANT OCCURRED ON (B)(6) 2014. IT IS UNKNOWN IF THIS WAS INTRA-OPERATIVE OR WHAT THE INITIAL IMPLANT DATE OF THE DEVICE IS.