FDA Adverse Event Malfunction Summary report: N

EMERGE?

MDR report key: 4193579 · Received October 22, 2014

Report

Report Number
2134265-2014-06389
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 30, 2014
Report Date
September 30, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
K113220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED AND MODERATELY TORTUOUS PROXIMAL LEFT ANTERIOR DESCENDING (LAD) ARTERY. A 1.20MM X 8MM EMERGE¿ BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE PHYSICIAN ATTEMPTED TO INCREASE PRESSURE LEVEL UP TO THE RATED BURST PRESSURE (RBP), HOWEVER, THE BALLOON RUPTURED AT AROUND 10 ATMOSPHERES. THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT¿S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
671826 EMERGE? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493919308120 16966726

Patients

Seq Age Sex Outcome Treatment
1