FDA Adverse Event Injury Summary report: N

STRATOS

MDR report key: 4193541 · Received September 11, 2014

Report

Report Number
3006801247-2014-00004
Event Type
Injury
Date Received
September 11, 2014
Report Date
September 9, 2014
Manufacturer
MEDXPERT GMBH
Product Code
HRS
PMA / PMN Number
K073556
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PECTUS EXCAVATUM (PE) TREATMENT WAS PERFORMED WITH ONE BRIDGE, CONSISTING OF ONE BAR AND TWO CLIPS. THE RECOMMENDATION FOR PE IS THE USE OF 3 BRIDGES. THE REOPERATION (EXPLANTATION) WAS PERFORMED 19 MONTHS AFTER IMPLANTATION. THE DECISION IF OR WHEN TO REMOVE IMPLANTS RESTS WITH THE SURGEON. IN ANY EVENT, POSTOPERATIVE EVALUATION OF THE PT IS MANDATORY. IT IS RECOMMENDED TO SEE EACH PT IN REASONABLE INTERVALS (EVERY 4-6 MONTHS). MATERIAL WAS NOT AVAILABLE DURING THIS REPORT FOR AN INSPECTION. LOT NUMBERS ARE NOT AVAILABLE FOR REPORTING ABOUT PRODUCTION AND MATERIAL DETAILS.

Description of Event or Problem · 1

DEFORMITY TREATMENT PECTUS EXCAVATUM, DETAILS (HALLER INDEX, SYMMETRIC/ASYMMETRIC) ARE NOT AVAILABLE. PT = FEMALE, BORN (B)(6) 1991. OPERATION DATE: (B)(6) 2012. THE 2 STRATOS CLIPS REF 012-02450, BOTH BROKEN IN THE ANGLE DATE OF BREAKAGE NOT DETECTABLE, POST-OPERATIVE PROBLEMS/OBSERVATIONS: WOUND PAIN. EXPLANTATION (B)(6) 2014. PRODUCTS AVAILABLE FOR INSPECTION: NO INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
563323 STRATOS TITANIUM RIB CLIPS HRS MEDXPERT GMBH 012-02450 UNK

Patients

Seq Age Sex Outcome Treatment
1