FDA Adverse Event Injury Summary report: N

STRATOS

MDR report key: 4193540 · Received September 11, 2014

Report

Report Number
3006801247-2014-00003
Event Type
Injury
Date Received
September 11, 2014
Report Date
September 9, 2014
Manufacturer
MEDXPERT GMBH
Product Code
HRS
PMA / PMN Number
K073556
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PECTUS CARINATUM (PC) TREATMENT WAS PERFORMED WITH ONE BRIDGE, CONSISTING OF ONE BAR AND TWO CLIPS. THE RECOMMENDATION FOR PC IS THE USE OF 3 BRIDGES. THE REOPERATION (EXPLANATION) WAS PERFORMED 21 MONTHS AFTER IMPLANTATION. THE DECISION IF OR WHEN TO REMOVE IMPLANTS RESTS WITH THE SURGEON. IN ANY EVENT, POSTOPERATIVE EVALUATION OF THE PATIENT IS MANDATORY. IT IS RECOMMENDED TO SEE EACH PATIENT IN REASONABLE INTERVALS (EVERY 4-6 MONTHS). MATERIAL WAS NOT AVAILABLE DURING THIS REPORT FOR AN INSPECTION.

Description of Event or Problem · 1

DEFORMITY TREATMENT PECTUS CARINATUM, MORE DETAILS (SYMMETRIC/ASYMMETRIC) ARE NOT AVAILABLE. PATIENT = FEMALE, BORN 1998 (B)(6). OPERATION DATE: 2012 (B)(6). STRATOS CLIPS REF 012-01225, 012-02225, STRATOS BAR 014-10190: THE BAR SLIDED OUT OF THE LUGS OF THE CLIPS, DATE OF INCIDENT NOT DETECTABLE, EXPLANTATION 2014 (B)(6). PRODUCTS AVAILABLE FOR INSPECTION: NO INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
563239 STRATOS TITANIUM CONNECTING BAR, CLIPS HRS MEDXPERT GMBH

Patients

Seq Age Sex Outcome Treatment
1