FDA Adverse Event Injury Summary report: N

STRATOS

MDR report key: 4193538 · Received September 11, 2014

Report

Report Number
3006801247-2014-00001
Event Type
Injury
Date Received
September 11, 2014
Report Date
September 9, 2014
Manufacturer
MEDXPERT GMBH
Product Code
HRS
PMA / PMN Number
K073556
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PECTUS CARINATUM (PC) TREATMENT WAS PERFORMED WITH ONE BRIDGE, CONSISTING OF ONE BAR AND TWO CLIPS. THE RECOMMENDATION FOR PE IS THE USE OF 3 BRIDGES. THE REOPERATION (EXPLANTATION) WAS PERFORMED 17 MONTHS AFTER IMPLANTATION. THE DECISION IF OR WHEN TO REMOVE IMPLANTS RESTS WITH THE SURGEON. IN ANY EVENT, POSTOPERATIVE EVALUATION OF THE PATIENT IS MANDATORY. IT IS RECOMMENDED TO SEE EACH PATIENT IN REASONABLE INTERVALS (EVERY 4-6 MONTHS). MATERIAL WAS NOT AVAILABLE DURING THIS REPORT FOR AN INSPECTION. LOT NUMBERS ARE NOT AVAILABLE FOR REPORTING ABOUT PRODUCTION AND MATERIAL DETAILS.

Description of Event or Problem · 1

DEFORMITY TREATMENT PECTUS EXCAVATUM, DETAILS (HALLER INDEX, SYMMETRIC/ASYMMETRIC) ARE NOT AVAILABLE. PATIENT = MALE, BORN (B)(6) 1941. OPERATION DATE: (B)(6) 2013. STRATOS BAR REF 014-10190 BROKEN ATTACHED TO THE LUG OF LATERAL CLIP. DATE OF BREAKAGE NOT DETECTABLE, EXPLANTATION (B)(6) 2014. PRODUCTS AVAILABLE FOR INSPECTION: FOR THIS REPORT NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
563254 STRATOS TITANIUM CONNECTING BAR HRS MEDXPERT GMBH 014-10190 UNK

Patients

Seq Age Sex Outcome Treatment
1