EON MINI
Report
- Report Number
- 1627487-2014-24297
- Event Type
- Injury
- Date Received
- October 20, 2014
- Date of Event
- September 26, 2014
- Report Date
- September 26, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
RECALL: 1627487-12192011-003-R. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORIES AND FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AN IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2014-24298. THE PATIENT REPORTED EXPERIENCING POCKET HEATING WHILE CHARGING. AS RESULT, A REPLACEMENT CHARGER WAS SENT TO THE PATIENT AND RESOLVED THE ISSUE. ON (B)(6) 2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 667159 | EON MINI | SCS IPG | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3224309 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Other | SCS LEAD: MODEL 3186 (2)| IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2) |