FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4193036 · Received October 22, 2014

Report

Report Number
2032227-2014-41767
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 23, 2014
Report Date
September 23, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE DOWN ARROW BUTTON DUE TO CORRODED KEYPAD TRACE. NO BUTTON ERROR ALARM NOTED. THE INSULIN PUMP HAS MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 72 MG/DL. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.