FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4193033 · Received October 22, 2014

Report

Report Number
2032227-2014-41779
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 23, 2014
Report Date
September 23, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP HAD A CRACKED CASE AT A DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD A BUTTON ERROR ALARM ON THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 6.9 MMOL/L. CUSTOMER STATED THAT IT MAY HAVE BEEN SQUASHED TOGETHER IN A CROWD. INSULIN PUMP WILL NEED TO BE REPLACED. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
671504 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554CAP

Patients

Seq Age Sex Outcome Treatment
1