LOANER REMB ELC WIREDRIVER
Report
- Report Number
- 0001811755-2014-03718
- Event Type
- Malfunction
- Date Received
- October 22, 2014
- Date of Event
- September 25, 2014
- Report Date
- September 25, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K112593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED. FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
THE REPORTED EVENT WAS NOT DUPLICATED; HOWEVER, UPON VISUAL INSPECTION, THE DIAGNOSTIC ENGINEER FOUND DELAMINATING FLEX. THE DEVICE WAS REPAIRED AND PLACED BACK INTO STRYKER STOCK.
IT WAS REPORTED THAT DURING SET-UP FOR A PROCEDURE AT THE USER FACILITY, THE REMB ELECTRIC WIREDRIVER WAS RUNNING WITHOUT USER ACTIVATION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO DELAY, NO ADVERSE CONSEQUENCES, AND MEDICAL INTERVENTION WERE REPORTED.
IT WAS REPORTED THAT DURING SET-UP FOR A PROCEDURE AT THE USER FACILITY, THE REMB ELECTRIC WIREDRIVER WAS RUNNING WITHOUT USER ACTIVATION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO DELAY, NO ADVERSE CONSEQUENCES, AND MEDICAL INTERVENTION WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 673500 | LOANER REMB ELC WIREDRIVER | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |