FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4192744 · Received October 22, 2014

Report

Report Number
2032227-2014-41568
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 23, 2014
Report Date
September 23, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND DUE TO CRACKED MOTOR FLEX CABLE. THE INSULIN PUMP WAS UNABLE TO CONFIRM ALARM IN ALARM HISTORY DUE TO MOTOR ERROR ALARMS. THE INSULIN PUMP HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON DISPLAY WINDOW NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING BASAL RATE DELIVERY. THE DEVICE WAS NOT EXPOSED TO A MAGNETIC FIELD. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND IS NOT ABLE TO COMPLETE THE REWIND SEQUENCE. IT WAS STATED THAT THEN A MOTOR POSITION ENCODER ERROR ALARM OCCURRED WHEN TRYING TO REWIND THE INSULIN PUMP. BLOOD GLUCOSE VALUE IS 292 MG/DL; TREATED WITH MANUAL INSULIN INJECTION. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673020 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAS

Patients

Seq Age Sex Outcome Treatment
1 11 YR