GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Report
- Report Number
- 2024168-2014-06887
- Event Type
- Injury
- Date Received
- October 22, 2014
- Date of Event
- September 29, 2014
- Report Date
- September 29, 2014
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MAF
- PMA / PMN Number
- H000001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY/REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. THE ADDITIONAL 4.0 X 18 MM XIENCE XPEDITION AND 4.0 X 16 MM GRAFTMASTER REFERENCED ARE BEING FILED UNDER SEPARATE MEDWATCH REPORTS.
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LESION IN THE MID RIGHT CORONARY ARTERY. THE 4.0 X 18 MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LESION AND DEPLOYED. AFTER DEPLOYMENT, A DISTAL DISSECTION WAS OBSERVED. TWO NON-ABBOTT STENTS WERE PLACED TO TREAT THE DISSECTION, BUT IT WAS THEN NOTED THAT THE INITIAL DISTAL DISSECTION BECAME A PERFORATION. THE 3.5 X 16 MM GRAFTMASTER STENT WAS DEPLOYED FOR TREATMENT, BUT THE PERFORATION WAS NOT SEALED. THE 4.0 X 16 MM GRAFTMASTER STENT WAS ADVANCED, BUT FAILED TO CROSS DUE TO THE ANATOMY. A 2.8 X 16 MM GRAFTMASTER WAS THEN ADVANCED AND DEPLOYED, AND THE PERFORATION WAS SEALED SUCCESSFULLY. THE PATIENT HAD A GOOD OUTCOME. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 671176 | GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM | CORONARY STENT GRAFT | MAF | AV-TEMECULA-CT | 3061241 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Required Intervention | STENT: 4.0 X 18 MM XIENCE XPEDITION |