FDA Adverse Event Malfunction Summary report: N

SPINEMAP® 3D 2.0 SOFTWARE

MDR report key: 4192687 · Received October 22, 2014

Report

Report Number
0001811755-2014-03714
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 25, 2014
Report Date
September 25, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HAW
PMA / PMN Number
K012380
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT WAS CONFIRMED DURING THE REVIEW OF LOG FILES AND PATIENT. THE REVIEW INDICATED THAT THE EVENT WAS DUE TO A DIFFERENT SPINE POSITION DURING REGISTRATION AGAINST THE PRE-OPERATIVE SCAN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SPINAL FUSION PROCEDURE OF T12 - L3 USING POINT-TO-POINT REGISTRATION WITH NAVIGATION SYSTEM II, THE SURGEON STATED THAT THE UPPER LEVELS AROUND T11 - L2 WERE INACCURATE FROM A LATERAL PERSPECTIVE. THE REGISTRATION WAS NEVER ACCEPTED AFTER TWO ATTEMPTS AT SURFACE REFINING. AS THE ISSUE WAS NOTICED PRIOR TO INSERTION OF VERTEBRAL SCREWS, NAVIGATION WAS ABORTED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING TRADITIONAL METHODS. NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SPINAL FUSION PROCEDURE OF T12 - L3 USING POINT-TO-POINT REGISTRATION WITH NAVIGATION SYSTEM II, THE SURGEON STATED THAT THE UPPER LEVELS AROUND T11 - L2 WERE INACCURATE FROM A LATERAL PERSPECTIVE. THE REGISTRATION WAS NEVER ACCEPTED AFTER TWO ATTEMPTS AT SURFACE REFINING. AS THE ISSUE WAS NOTICED PRIOR TO INSERTION OF VERTEBRAL SCREWS, NAVIGATION WAS ABORTED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING TRADITIONAL METHODS. NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
671091 SPINEMAP® 3D 2.0 SOFTWARE NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW STRYKER INSTRUMENTS-KALAMAZOO VERSION 2.0-22

Patients

Seq Age Sex Outcome Treatment
1