FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4192679 · Received October 22, 2014

Report

Report Number
2032227-2014-41101
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 22, 2014
Report Date
September 22, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD MOISTURE DAMAGE WAS FOUND ON THE KEYPAD TRACES. ALL BUTTONS RESPONDED PROPERLY. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, BROKEN BELT CLIP SLOT AND MINOR SCRATCHES ON LCD WINDOW.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP HAD A LOW PREDICT ALARM AND THE ESC BUTTON WAS UNRESPONSIVE. CUSTOMER STATED THAT THE INSULIN PUMP WAS IN THE BATHROOM WITH HER WHILE TAKING A SHOWER. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 88 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673328 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 51 YR