FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4192642 · Received October 22, 2014

Report

Report Number
2032227-2014-41096
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 22, 2014
Report Date
September 22, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO NUMBER RAMPING ANOMALY WAS NOTED ON THE DISPLAY. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE NUMBERS ON THE INSULIN PUMP WERE RAMPING UP. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 171 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
672157 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAB

Patients

Seq Age Sex Outcome Treatment
1 54 YR