FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4192617 · Received October 22, 2014

Report

Report Number
2032227-2014-41488
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
September 2, 2014
Report Date
September 22, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST AND EXCESSIVE NO DELIVERY ALARM TEST. NO UNEXPECTED EXCESSIVE NO DELIVERY ALARM WAS NOTED. THE INSULIN WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A SCRATCHED RESERVOIR TUBE WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTS TO HAVE RECEIVED EXCESSIVE DELIVERY ALARMS. CUSTOMER'S BLOOD GLUCOSE WAS 168 MG/DL. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
672452 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAB

Patients

Seq Age Sex Outcome Treatment
1 68 YR