PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-41496
- Event Type
- Malfunction
- Date Received
- October 22, 2014
- Date of Event
- September 23, 2014
- Report Date
- September 23, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP RECEIVED WITH UNRESPONSIVE BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS AND BELT CLIP SLOT.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4).
NO BLOOD GLUCOSE LEVELS WERE PROVIDED. THE CUSTOMER REPORTED A LOW RESERVOIR ALARM FROM THE INSULIN PUMP THAT WAS NOT CLEARED BECAUSE SOME OF THE BUTTONS OF THE INSULIN PUMP WERE UNRESPONSIVE. THE CUSTOMER ALSO REPORTED THAT THE LIGHT ON THE INSULIN PUMP WAS TURNED ON. THE CUSTOMER FURTHER REPORTED A BUTTON ERROR ALARM FROM THE INSULIN PUMP. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 671304 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR |