FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4192299 · Received October 22, 2014

Report

Report Number
2531779-2014-29941
Event Type
Malfunction
Date Received
October 22, 2014
Report Date
October 9, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 12/03/2014 WITH THE FOLLOWING FINDINGS: A 'LOSS OF PRIME' WARNING, WHICH WAS DUE TO A NON-ZERO LOW FORCE READING AND CAN BE INDICATIVE OF THE TYPE OF ISSUE THAT WAS REPORTED, WAS OBSERVED IN THE BLACK BOX DATA. THE PUMP WAS EXERCISED FOR 24 HOURS, DURING WHICH NO 'LOSS OF PRIME' WARNINGS WERE OBSERVED: THE REPORTED ISSUE WAS NOT DUPLICATED. THE PUMP FAILED A FORCE SENSOR CALIBRATION CHECK DUE TO A FORCE SENSOR CALIBRATION READING THAT WAS BELOW THE LOWER SPECIFICATION LIMIT. THE PUMP CASE WAS REMOVED, AND NO EVIDENCE OF INTERNAL DAMAGE OR DEFECT WAS FOUND. THE PUMP PASSED A FORCE SENSOR RESISTANCE CHECK.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. IT WAS REPORTED THAT A LOSS OF PRIME WARNING OCCURRED 3 OR MORE TIMES WITHIN A 30 DAY PERIOD, DURING WHICH AT LEAST 3 SEPARATE CARTRIDGES FROM 2 SEPARATE BOXES WERE USED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673865 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 44 YR