FDA Adverse Event Malfunction Summary report: N

ISOFLEX OPTIM LEAD

MDR report key: 4192144 · Received January 13, 2014

Report

Report Number
2017865-2014-05481
Event Type
Malfunction
Date Received
January 13, 2014
Date of Event
January 23, 2013
Manufacturer
ST. JUDE MEDICAL, CRMD
Product Code
DTB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFO WAS PROVIDED BY MFR, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVAL DESCRIPTION: FINAL ANALYSIS CONFIRMED INSULATION DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED LOW IMPEDANCE AT IMPLANT. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
30301 ISOFLEX OPTIM LEAD PERMANENT PACEMAKER ELECTRODE DTB ST. JUDE MEDICAL, CRMD 1948/52

Patients

Seq Age Sex Outcome Treatment
1 68 YR